Related Experiment Video
Updated: Aug 7, 2026

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Precise ulna-shortening osteotomy with a new device
1Division of Orthopedic Surgery, Hiroshima Prefectural Rehabilitation Center, Higashi-Hiroshima City, Japan.
Abstract:
Ulna-shortening osteotomy has become a standard procedure for various ulnar-sided wrist disorders. A precise osteotomy, good coaptation of the osteotomy surfaces, and rigid internal fixation are mandatory to achieve good results from ulna-shortening osteotomy. Various techniques and devices have been introduced to assist in this difficult procedure. We developed a device that enables a precise ulna-shortening osteotomy and fixation with a 3.5-mm AO dynamic compression plate. Twenty-four ulnas were shortened with the aid of this device. The average follow-up period was 36 months. Radiologic union occurred at an average of 8.1 weeks after surgery. There were no nonunions or delayed unions. This study shows that an ulna-shortening procedure with this device enables precise shortening and predictable union of the ulna.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
07:33In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023