Related Experiment Videos
[Excimer laser extraction of pacemaker and defibrillator leads]
Summary
Excimer laser lead extraction offers a high success rate for removing pacemaker and defibrillator leads, preventing complications from retained fragments. While effective, serious complications can occur with both laser and conventional methods, necessitating specialized centers for patient safety.
Area of Science:
- Cardiology
- Medical Devices
- Surgical Technology
Background:
- Pacemaker or defibrillator lead malfunction/infection necessitates foreign material removal.
- Laser energy offers a novel approach for extracting long-dwelling, fixed leads.
- Limited comparative data exists on laser extraction versus conventional methods.
Purpose of the Study:
- To compare the efficacy and safety of excimer laser lead extraction with conventional methods.
- To evaluate complication rates and success rates for different lead extraction techniques.
Main Methods:
- Utilized Spectranetics excimer laser system (CVX-300) with laser sheaths and a lead locking device.
- Compared laser extraction in 24 patients (45 leads) with conventional methods (manual traction, traction devices, snare catheters, thoracotomy) from 1995-1998 (97 leads).
- Procedures performed under heart-lung machine backup.
Main Results:
- Laser extraction achieved a 96% success rate with minimal fragmentation; conventional methods had a 76% success rate, with fragments remaining in 15% of cases.
- Laser method operation time (93 min) and fluoroscopy time (9.4 min) were comparable to conventional techniques.
- One fatal complication (cerebral hypoxia) occurred with laser extraction; conventional methods resulted in one fatality (sepsis) and four severe complications.
Conclusions:
- Excimer laser extraction facilitates complete removal of leads, preventing late complications from retained fragments.
- Life-threatening complications are possible with both laser and conventional lead extraction.
- Lead extraction procedures should be conducted in specialized centers with advanced monitoring capabilities.