Comparison of patient dose from imaging protocols for dental implant planning using conventional radiography and
A R Lecomber1, Y Yoneyama, D J Lovelock
1Regional Medical Physics Department, Royal Victoria Infirmary, Newcastle upon Tyne, UK.
Objectives:
To compare the radiation doses from imaging protocols for dental implant planning either using conventional radiography only (dental panoramic radiography (DPR), cephalometry and linear cross-sectional tomography) or involving computed tomography (CT).
Methods:
Organ absorbed doses were measured using a female Rando anthropomorphic phantom loaded with lithium fluoride thermoluminescent dosemeters (TLD). Standard mandibular protocols for dental implant planning were followed using either a conventional dental radiographic unit (PM 2002 CC Planmeca, Helsinki, Finland) or CT scanner (Excel Twin Elscint, Haifa, Israel). Organ absorbed and effective doses were calculated. Effective dose was calculated using two approaches, one based on the ICRP method which excludes the salivary tissue from the remainder organs (designated E(exc)), and the other with its inclusion (E(inc)).
Results:
The greatest individual organ doses for any examination were measured in the salivary tissue. E(exc) for panoramic, cephalometric and cross-sectional tomography using DPR was 0.004 mSv, 0.002 mSv and 0.002 mSv, respectively, whereas with CT it was 0.314 mSv. The value of E(inc) calculated using these data was between two and five times E(exc).
Conclusions:
E(inc) greatly increases the apparent radiation burden, especially with high dose procedures. CT techniques can provide excellent images, but at the cost of increased radiation detriment. DPR with a cross-sectional tomography facility may give adequate clinical information at a greatly reduced dose.
More Related Videos
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography