Related Experiment Video
Updated: Jul 12, 2026

Assessment of Kidney Function in Mouse Models of Glomerular Disease
Published on: June 30, 2018
Albuminuria and proteinuria in hospitalized patients as measured by quantitative and dipstick methods
M J Pugia1, J F Wallace, J A Lott
1Diagnostics Business Group, Bayer Corporation, Elkhart, Indiana, USA.
Abstract:
We tested patients' urines for albumin, protein, and creatinine by quantitative and dipstick methods. The concentrations of these analytes were established by quantitative, cuvet-based chemistry methods that we assumed gave the "correct" values. There was good to excellent agreement of the dipstick results with the quantitative methods for the above three analytes. We found many patients who excreted pathological amounts of albumin and/or protein who did not have a diagnosis of kidney disease or other likely causes of proteinuria, suggesting that albuminuria and/or proteinuria were underdiagnosed in our group of patients. Those with cardiovascular disease, kidney disease, or diabetes showed the greatest predictive value of a positive test for albumin or protein by dipstick. Dipstick testing for albumin, protein, and creatinine had good or excellent agreement with quantitative methods. The dipstick tests were easy to use, simple, and low in cost, and can serve well for point-of-care testing.
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Nephrotic Syndrome II : Assessment and Medical Management
Urine Studies I: Urinalysis
Serum Studies: Renal Function Tests
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Diabetic Nephropathy