Frameless optical computer-aided tracking of a microscope for otorhinology and skull base surgery
G Zheng1, M Caversaccio, R Bächler
1Department of Oto-Rhino-Laryngology, Head and Neck Surgery, University Hospital, Freiburgstrasse, CH-3010 Bern, Switzerland.
Objectives:
To integrate a digitally controlled operating microscope without a laser autofocus system into a frameless optical computer-aided surgery system and to test the accuracy and usability of this system in otorhinological surgery.
Design:
Experimental study and case series.
Setting:
Department of Oto-Rhino-Laryngology, Head and Neck Surgery, Inselspital, and the Maurice E. Müller Institute for Biomechanics, University of Bern, Bern, Switzerland.
Patients:
Eight computer-aided microscopic surgical procedures were performed between January and October 2000 on patients with various diseases of the anterior and lateral skull base.
Results:
The practical accuracy of the navigated microscope on the lateral side of a cadaver skull was 2.27 +/- 0.25 mm and on the anterior side of the same skull was 2.07 +/- 0.35 mm. In all 8 cases of computer-aided microscopic surgery, no complications occurred. Clinical inaccuracy was 2 to 3 mm.
Conclusion:
Integration of a low-cost, non-laser autofocus microscope into our computer-aided surgery system was successfully performed and offers surgeons the ability to combine the precise optics of the operating microscope with the localization power of a computer-aided system.