In vitro performance characteristics of high-frequency oscillatory ventilators
J J Pillow1, M H Wilkinson, H L Neil
1TVW Telethon Institute for Child Health Research, Perth, Australia. jpillow@binternet.com
Abstract:
This study aimed to examine the performance characteristics of four high-frequency oscillatory-type ventilators, using an in vitro model of the intubated neonatal respiratory system. Each ventilator was examined across its operative range of settings and at varying model lung compliance (C) and resistance. The oscillatory pressure waveform was measured at the airway opening (Pao). Tidal volume (VT) and flow were determined from pressure changes within the model lung (DeltaPA). The spectral content of the Pao waveform differed between ventilators. The maximum ventilator VT ranged from 3.7 to 11.1 ml at 15 Hz and a mean airway pressure (Paw) of 12 cm H(2)O to oscillate a model lung (C = 0.4 ml/cm H(2)O) through a 3.0-mm internal diameter (i.d.) endotracheal tube (ETT). A small drop in C was associated with a decrease in VT and marked increase in DeltaPA from 0.1 to 0.8 ml/cm H(2)O. The influence of C on VT and DeltaPA and the pressure cost of ventilation (DeltaPA/f.VT(2)) was dependent on the oscillatory frequency, ETT inner diameter, and the specific ventilator used. Substantive differences exist between oscillatory ventilators that need to be considered in their clinical application. The rapid establishment of optimal lung volume and oscillatory frequency is important in minimizing barotrauma during high-frequency oscillatory ventilation.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...