[Ocular biometric measurements with a slit-lamp method (Orbscan)].

C Allouch1, O Touzeau, V Borderie

  • 1Pôle Saint Antoine/CHNO des Quinze-Vingts, 28, rue de Charenton, 75571 Paris Cedex 12, France.

Journal Francais D'Ophtalmologie
|October 10, 2001
PubMed
Summary

The Orbscan((R)) device offers accurate anterior chamber depth and corneal diameter measurements in normal eyes, showing high reproducibility. However, axial length measurements with Orbscan((R)) were less reliable and not clinically applicable.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Efficacy and safety of intraocular folding sutureless scleral fixating lens versus iris-claw intraocular lens implantation.

Journal francais d'ophtalmologie·2022
Same author

Predictibility of the efficacy of peripheral iridotomy in primary angle closure: Utility of UBM.

Journal francais d'ophtalmologie·2022
Same author

[Irido-ciliary melanoma].

Journal francais d'ophtalmologie·2021
Same author

Epidemiological changes in cutaneous lymphomas: an analysis of 8593 patients from the French Cutaneous Lymphoma Registry.

The British journal of dermatology·2020
Same author

Boosting the sterile insect technique with pyriproxyfen increases tsetse flies Glossina palpalis gambiensis sterilization in controlled conditions.

Scientific reports·2020
Same author

Atypical severe diffuse lamellar keratitis presenting as concentric rings after femtosecond laser-assisted small-incision lenticule extraction (SMILE).

Journal francais d'ophtalmologie·2020

Area of Science:

  • Ophthalmology
  • Biomedical Optics
  • Medical Imaging

Background:

  • Accurate biometric measurements are crucial for ophthalmic diagnostics and surgical planning.
  • The Orbscan((R)) is a device used for anterior segment analysis.
  • Evaluating device accuracy and reproducibility is essential for clinical adoption.

Purpose of the Study:

  • To assess the accuracy and reproducibility of biometric measurements using the Orbscan((R)) device.
  • To compare Orbscan((R)) measurements with A-Scan ultrasonography in normal eyes.

Main Methods:

  • Biometric measurements (anterior chamber depth, axial length, corneal diameter) were taken on 50 normal eyes using Orbscan((R)) and A-Scan ultrasonography.
  • Reproducibility was determined by analyzing the difference between two successive measurements.

Related Experiment Videos

  • Correlation analysis was performed between Orbscan((R)) and A-Scan ultrasonography measurements.
  • Main Results:

    • Orbscan((R)) demonstrated high reproducibility for anterior chamber depth (0.72±1.25%) and corneal diameter (0.83±1.40%) measurements.
    • Anterior chamber depth measurements showed strong correlation between Orbscan((R)) and A-Scan ultrasonography (rs=0.92, p<0.001).
    • Axial length measurements with Orbscan((R)) exhibited lower reproducibility (3.55±4.11%) and poor correlation with A-Scan ultrasonography (rs=0.64, p<0.001).

    Conclusions:

    • Orbscan((R)) provides highly reproducible and accurate anterior chamber depth measurements, correlating well with A-Scan ultrasonography.
    • Axial length measurements using Orbscan((R)) are less reproducible and poorly correlated with A-Scan ultrasonography, limiting clinical relevance.
    • The Orbscan((R)) device is suitable for anterior chamber depth and corneal diameter assessment but not recommended for axial length measurements in clinical practice.