A comparison of steroid-eluting epicardial versus transvenous pacing leads in children

A Dodge-Khatami1, C L Johnsrude, C L Backer

  • 1Department of Surgery, Children's Memorial Hospital, Northwestern University Medical School, Chicago, Illinois 60614, USA.

Insights

Steroid-eluting epicardial (EPI) pacing leads show performance equivalent to transvenous (TV) leads in children. EPI leads are recommended for long-term pacing when feasible.

Area of Science:

  • Pediatric Cardiology
  • Biomedical Engineering
  • Electrophysiology

Background:

  • Congenital heart disease often necessitates long-term pacing in children.
  • Steroid-eluting (SE) pacing leads aim to improve lead performance and longevity.
  • Epicardial (EPI) and transvenous (TV) lead implantation techniques have distinct characteristics.

Purpose of the Study:

  • To compare the acute and chronic performance of SE epicardial (EPI) pacing leads versus SE transvenous (TV) pacing leads in pediatric patients.
  • To assess sensing thresholds, capture thresholds, and impedance for both lead types.

Main Methods:

  • A retrospective analysis of 55 children with congenital heart disease who received 85 SE pacing leads (38 EPI, 47 TV) between 1989 and 1997.
  • Evaluation of acute and chronic sensing and capture thresholds, and impedance for all implanted leads.
  • Comparison of patient demographics, including age at implantation and follow-up duration.

Main Results:

  • Acute and chronic sensing and capture thresholds were comparable between SE EPI and SE TV leads.
  • Steroid-eluting transvenous (TV) leads demonstrated consistently higher chronic impedance compared to steroid-eluting epicardial (EPI) leads.
  • Patients receiving EPI leads were younger at implantation and had shorter follow-up periods.

Conclusions:

  • Steroid-eluting epicardial (EPI) pacing leads are functionally equivalent to steroid-eluting transvenous (TV) leads at intermediate follow-up.
  • The study recommends considering SE EPI leads before resorting to the TV approach for children requiring lifelong pacing.
  • This finding supports the use of EPI leads as a viable option in pediatric cardiac pacing.