A comparison of prospective and retrospective respiratory navigator gating in 3D MR coronary angiography
Y P Du1, E R McVeigh, D A Bluemke
1Applied Science Lab, GE-Medical Systems, Milwaukee, Wisconsin, USA. yipingdu@uchicago.edu
Abstract:
A comparison between the prospective and retrospective respiratory navigator gating in MR coronary angiography was performed with eight normal subjects. A three-dimensional (3D) ECG-gated fast gradient echo pulse sequence was used for image data acquisition. The results show that the MR coronary angiography obtained using retrospective gating retains a considerable amount of motion artifacts. In this study, the images acquired using prospective navigator gating demonstrated significantly reduced motion artifacts (p = 0.009), improved vessel visibility (p = 0.021) with reduced imaging time (p = 0.013) compared to the images obtained using retrospective navigator gating.