Related Experiment Video
Updated: Jul 31, 2026

Clean Sampling and Analysis of River and Estuarine Waters for Trace Metal Studies
Published on: July 1, 2016
Extraction and analysis of lead in sweeteners by flow-injection Donnan dialysis with flame atomic absorption
1Department of Chemistry, University of Wisconsin-Parkside, 900 Wood Road, Kenosha, Wisconsin 53141, USA.
Abstract:
Flow-injection Donnan dialysis is demonstrated for the extraction of lead in sweeteners using flame atomic absorption spectroscopy (FAAS). For spiked concentrations in the low microgram per gram range, recoveries were greater than 90%, and the relative standard deviation was typically less than 10% for a 15-min dialysis procedure. The method detection limit is 350 ng/g. Donnan dialysis is shown to be successful for the extraction of lead in sucrose, corn syrup, and honey but limited in performance for molasses and artificial syrup. This paper also includes a comparison to other procedures for the determination of lead in sweeteners and presents options for realizing improved method performance with Donnan dialysis.
More Related Videos
Related Concept Videos
Washing, Drying, and Ignition of Precipitates
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...