Use of infrared emission detection thermometer in Chinese neonates
1Department of Pediatrics, Kwong Wah Hospital, 25 Waterloo Road, Kowloon, Hong Kong SAR, China.
Objective:
To evaluate the reproducibility of Thermoscan, an infrared emission detection ear thermometer, and to establish the normal reference range of ear temperature in Chinese neonates.
Methods:
Neonates were recruited from the inpatients population with exclusion of those suffering from infections.
Results:
Forty-nine neonates were recruited with 1,115 temperature taking sessions. Mean left ear temperature was 36.64 degrees C +/- 0.35 degree C. Mean right ear temperature was 36.64 degrees C +/- 0.37 degree C. Clinical repeatability for left and right ear was 0.17 degree C and 0.17 degree C, respectively.
Conclusion:
Thermoscan produced reproducible results in Chinese neonates. The normal range of ear temperature for Chinese neonates is 35 degrees C to 37 degrees C. Ear temperature > 37.8 degrees C should be regarded as fever.