[Reproducibility and interchangeability of the Thromboelastograph, Sonoclot and Hemochron activated coagulation time

F Forestier1, S Bélisle, C Contant

  • 1Département D'anesthésiologie, Institut de Cardiologie de Montréal, Montréal, Québec, Canada.

Insights

The Thromboelastograph (TEG) offers the most reproducible on-site hemostasis monitoring in cardiac surgery. Other tests like Sonoclot (SCT) and Hemochron are less reproducible, and results are not interchangeable in heparinized patients during cardiopulmonary bypass.

Area of Science:

  • Cardiovascular Surgery
  • Hemostasis Monitoring
  • Point-of-Care Testing

Background:

  • On-site hemostasis monitoring is crucial in cardiac surgery.
  • Reproducibility of Thromboelastograph (TEG), Sonoclot (SCT), and Activated Coagulation Time (ACT) devices is not well-established in clinical settings.

Purpose of the Study:

  • To determine the reproducibility of TEG, SCT, and Hemochron ACT measurements (ACT-C, ACT-K) during on-site hemostasis monitoring.
  • To evaluate the interchangeability of ACT-K and SCT with ACT-C under various clinical conditions.

Main Methods:

  • Blood samples from 20 volunteers and 21 cardiac surgery patients were analyzed.
  • TEG, SCT, and Hemochron analyzers were used.
  • Variation rate (V%) was calculated for each parameter; interchangeability was assessed.

Main Results:

  • TEG demonstrated the highest reproducibility (3.1-9.5% V%), followed by Hemochron (5.6-12.4% V%) and SCT (5.8-33.6% V%).
  • ACT-C and ACT-K were interchangeable in non-heparinized patients.
  • No other test could substitute for ACT-C in heparinized patients during cardiopulmonary bypass (CPB).

Conclusions:

  • TEG provides the most reproducible on-site hemostasis measurements, followed by Hemochron and SCT.
  • Instrument-specific values are essential for monitoring anticoagulation in heparinized patients during CPB due to non-interchangeable results.