Sphygmomanometers in use in general practice: an overlooked aspect of quality in patient care
1Directorate of Health Promotion, North Staffordshire Health, 79 London Road, Stoke-on-Trent ST4 7PZ, UK. T.M.Knight@BHAM.ac.uk
Objectives:
To assess the condition, accuracy and safety of mercury and anaeroid sphygmomanometers in use in general practice and to pilot a scheme for sphygmomanometer maintenance within the district.
Design:
Instruments were checked on site according to set protocols which related to established guidelines and standards and data were entered into a specially designed database. Mercury sphygmomanometers were removed to the laboratory for servicing. Practices received written feedback on the condition of each instrument checked, repairs undertaken and advice, where necessary, for further work required. Participant views on the scheme were sought.
Materials And Methods:
A total of 472 instruments (75.4% mercury) located in 86 general practices (87.8% of practices participated) in a health district in the West Midlands. Instruments were checked against 28 (mercury) and 25 (anaeroid) quality standards and (for mercury instruments) against British Hypertension Society guidelines.
Results:
69.1% of mercury and 95.7% of anaeroid instruments checked, had no service records. Of the remainder, only 29 mercury (8.1%) and one anaeroid (0.9%) had a record of a check or service within the previous 12 months. None of the instruments met all of the relevant quality standards and 14 (3.9%) mercury and seven (6.1%) anaeroid instruments met less than half. A large proportion of mercury sphygmomanometers tested had defects likely to affect recommended measurement technique. Only two-thirds were accurate at all pressure levels tested. Only 38.8% of anaeroid instruments were accurate at all test pressure levels.
Conclusions:
The level of defects noted could have an impact on diagnosis and monitoring of hypertension.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: