Accuracy of three-dimensional quantification of left ventricular function using magnetic sensor acquisition: a

J Xu1, Y Wu, X Li

  • 1Department of Ultrasound, Shenzhen People's Hospital, Shenzhen 518020, China. xujinf@163.net

Chinese Medical Journal
|October 27, 2001
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Quantification of the Mg2+-induced potency shift of amantadine and memantine voltage-dependent block in human recombinant GluN1/GluN2A NMDARs.

Neuropharmacology·2010
Same author

Investigation into the efficacy of the acetylcholinesterase inhibitor, donepezil, and novel procognitive agents to induce gamma oscillations in rat hippocampal slices.

Neuropharmacology·2010
Same author

Multifunctional catheters combining intracardiac ultrasound imaging and electrophysiology sensing.

IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2008
Same author

Accuracy of real-time, three-dimensional Doppler echocardiography for stroke volume estimation compared with phase-encoded MRI: an in vivo study.

Heart (British Cardiac Society)·2008
Same author

Semi-automatic segmentation of fetal cardiac cavities: progress towards an automated fetal echocardiogram.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2008
Same author

Characterization of a CNS penetrant, selective M1 muscarinic receptor agonist, 77-LH-28-1.

British journal of pharmacology·2008

Three-dimensional (3D) magnetic Flock of Birds (FOB) scanning accurately quantifies ventricular stroke volume in a dynamic model. This method shows improved correlation compared to 2D techniques, facilitating left ventricular function assessment.

Area of Science:

  • Medical Imaging
  • Cardiovascular Technology
  • Biomedical Engineering

Background:

  • Accurate quantification of ventricular stroke volume is crucial for assessing left ventricular function.
  • Traditional two-dimensional (2D) echocardiography methods have limitations in precise volume measurement.
  • Advancements in three-dimensional (3D) imaging technologies offer potential for improved accuracy.

Purpose of the Study:

  • To evaluate the accuracy of a 3D magnetic position sensor system (Flock of Birds - FOB) for quantifying ventricular stroke volumes.
  • To compare the accuracy of 3D FOB system with conventional 2D biplane methods.
  • To assess the practicality and utility of 3D FOB scanning in dynamic models.

Main Methods:

  • A latex balloon model simulating the left ventricle was used with controlled pulsatile stroke volumes (15-65 ml/beat).

Related Experiment Videos

  • Scanning was performed using a 5.0 mHz transducer with an integrated FOB magnetic receiver.
  • 3D Echopac software (GE) and a 2D biplane area-length method were used for stroke volume calculations and compared to direct measurements.
  • Main Results:

    • The 3D FOB system demonstrated a strong correlation with reference stroke volumes (r = 0.97, P = 0.0001).
    • This correlation was significantly improved compared to the 2D biplane method (r = 0.87, P = 0.0011).
    • The 3D method showed a lower standard error of the estimate (SEE = 2.83 ml) than the 2D method (SEE = 3.87 ml).

    Conclusions:

    • Three-dimensional magnetic FOB scanning is a practical and accurate method for quantifying ventricular stroke volumes.
    • The 3D FOB system offers superior accuracy over 2D methods for stroke volume quantification.
    • This technology should facilitate improved assessment of left ventricular function.