Gas chromatographic-mass spectrometric method for quantitation of phenylalanine and tyrosine in serum
J B Laurens1, X Y Mbianda, J B Ubbink
1Department of Chemical Pathology, Faculty of Medicine, University of Pretoria, South Africa. jlaurens@medic.up.ac.za
Abstract:
A validated gas chromatographic-mass spectrometric method for quantitation of phenylalanine and tyrosine in serum is described. Quantitation of phenylalanine and tyrosine with a non-labelled non-endogenous internal standard, L-2-chlorophenylalanine, compared favourably with isotope dilution mass spectrometric quantitation. The 95% reference ranges for phenylalanine. tyrosine and the phenylalanine-tyrosine molar ratio in neonate cord blood serum were determined by isotope dilution mass spectrometry and were found to be 77.1-144.7, 33.3-109.3 micromol/l and 1.1-3.0, respectively.
More Related Videos
10:30A Generalized Method for Determining Free Soluble Phenolic Acid Composition and Antioxidant Capacity of Cereals and Legumes
Published on: June 10, 2022
14:39Semi-Targeted Ultra-High-Performance Chromatography Coupled to Mass Spectrometry Analysis of Phenolic Metabolites in Plasma of Elderly Adults
Published on: April 22, 2022