Related Experiment Videos
Recognition of multiple classes of hepatitis C antibodies increases detection sensitivity in oral fluid
J F Zmuda1, B Wagoneer, L Liotta
1Immunomatrix Inc., Gaithersburg, Maryland 20878, USA. JonZmuda@clark.net
Clinical and Diagnostic Laboratory Immunology
|November 1, 2001
Abstract:
Paired serum-oral fluid samples from 127 hepatitis C virus (HCV)-positive and 31 HCV-negative patients were tested for the presence of anti-HCV using the Ortho HCV 3.0 ELISA. Using the immunoglobulin G (IgG)-specific detection antibody provided with the HCV 3.0 ELISA we attained 100% sensitivity and specificity with serum samples; however, sensitivity in oral fluid samples was only 81%. By modifying the HCV 3.0 ELISA to utilize a secondary antibody cocktail that recognizes not only IgG but IgA and IgM as well, we attained 100% specificity and sensitivity with oral fluid samples.