Breathing patterns and aerosol delivery: impact of regular human patterns, and sine and square waveforms on rate of
K Nikander1, J Denyer, N Smith
1Clinical Science, AstraZeneca R&D, Lund, Sweden. Kurt.nikander@astrazeneca.com
Abstract:
In vitro tests are commonly employed to assess nebulizer performance. Whether the square or sine waveforms employed during in vitro tests could alter the nebulizer performance compared to that observed when a patient breathes through the nebulizer is debatable. Accordingly, the aim of this in vitro study was to compare the rates of delivery from nebulizers with simulated human breathing patterns to those obtained with matching sine and square waveforms. Regular human breathing patterns with tidal volumes (VT) of approximately 40, approximately 200, approximately 500, and approximately 800 mL were selected. Sine and square waveforms that matched the VT, peak inspiratory flow rate (PIF), breathing frequency (f), and inspiratory duty cycle (t(i)/t(tot)) of the human breathing patterns were created with a breathing simulator. The rate of delivery of nebulized technetium-99m-labeled diethylenetriamine pentaacetic acid (99mTC-DTPA) from two different jet nebulizer brands was determined. The rate of delivery was defined as the amount of the 99mTC-DTPA deposited during 30 sec of nebulization on a filter placed between the nebulizer and the breathing simulator. The rate of delivery of 99mTC-DTPA with the human breathing pattern was similar to that measured with the matching sine or square waveforms for either nebulizer. The configuration of the breath (PIF, VT, f, t(i)/t(tot)) did, however, influence the rate of delivery. In conclusion, the shape of the waveform, in other words, one resulting from a human breathing pattern, or a matching sine or square waveform, did not influence the rate of 99mTC-DTPA delivery from a nebulizer in vitro.
Related Concept Videos
Respiratory Volumes and Capacities I
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes include...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...
Factors Affecting Respiration
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...