[Reproducibility of measurement of ocular pulse amplitude and intraocular pressure using Smartlens]
1Universitäts-Augenklinik, Langenbeckstrasse 1, 55131 Mainz. Annette.Vogel@gmx.de
Background:
SmartLens (ODC Ophthalmic Development Company AG, Zürich) is a contact lens tonometer, which allows continuous registration of intraocular pressure (IOP) and ocular pulse amplitude (OPA) and simultaneous ophthalmoscopy. The purpose of this study was to evaluate the intra- and interobserver variability and reliability of intraocular pressure and ocular pulse amplitude measurements using SmartLens.
Methods:
To evaluate the intraobserver variability, intraocular pressure (IOP) and ocular pulse amplitude (OPA) were both measured 5 times in 10 healthy eyes by one observer. To evaluate the interobserver variability, intraocular pressure and ocular pulse amplitude were measured by 3 different observers in 10 healthy eyes. Data description is based on coefficients of variation (SD/mean in %), statistical inference on ANOVA-based reliability estimation.
Results:
The coefficient of the intraobserver variation was 9.57% for the IOP and 14.5% for the OPA (SD IOP: +/- 1.64 mmHg, SD OPA: +/- 0.39 mmHg). The coefficient of the interobserver variation with 5 replicate measurements was 10.16% for the IOP and 14% for the OPA (SD IOP: +/- 1.83 mmHg, SD OPA: +/- 0.4 mmHg). The interobserver reliability was 87% for the IOP and 86% for the OPA. One of the three observers obtained significantly higher IOP and OPA values.
Conclusions:
SmartLens is a reasonably precise and reliable device for IOP as well as OPA measurements but the results were not found to be observer independent.
Related Concept Videos
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...