Development and validation of a high-performance liquid chromatographic method for the analysis of propylthiouracil
A M Abdul-Fattah1, H N Bhargava
1Department of Pharmaceutical Sciences, Massachusetts College of Pharmacy and Health Sciences, Boston, MA 02115, USA.
Abstract:
A simple, rapid, and stability-indicating high-performance liquid chromatographic (HPLC) method was developed and validated for the assay of propylthiouracil (PTU). The method was used to quantify PTU in topical formulations and in tablets. Excellent linearity was observed between PTU concentration and the peak area (R2= 0.999). The limit of detection was 1 ng, and the limit of quantitation was 1.2 ng. The method proved to be selective. Selectivity was validated by subjecting a stock solution of PTU to acidic, basic, and oxidative degradations. The peaks of the degradation products did not interfere with the peak of PTU. Excipients present in the dosage forms did not interfere with the analysis, and the recovery of PTU from each dosage form was quantitative.
More Related Videos
07:29HPLC Coupled with Chemical Fingerprinting for Multi-Pattern Recognition for Identifying the Authenticity of Clematidis Armandii Caulis
Published on: November 11, 2022
09:10An Efficient and Rapid HPLC-QQQ-MS Method for the Quantitation of Tropane Alkaloids in Medicinal Plants
Published on: June 12, 2026
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
Therapeutic Drug Monitoring: Drug Analysis Methods