Related Experiment Video
Updated: Jul 28, 2026

Programmed Electrical Stimulation in Mice
Published on: May 26, 2010
Does electrode polarity alter the energy requirements for transthoracic biphasic waveform defibrillation?
G Karlsson1, Y Zhang, L R Davies
1Department of Internal Medicine, The Cardiovascular Center, University of Iowa Hospitals and Clinics, 200 Hawkins Drive, Iowa City, IA 52242, USA.
Background:
Electrode polarity may alter the success of biphasic shocks from implantable systems. Whether the electrode polarity influences the success of transthoracic biphasic defibrillation is unknown. We determined the effect of electrode polarity on biphasic transthoracic defibrillation in a porcine model.
Method:
In ten anesthetized adult pigs, 16-28 kg, electrode pads were placed in two different orientations on the chest wall; apex-right parasternal and sternal-vertebral column. Ventricular fibrillation (VF) was electrically induced and allowed to persist for 30 s. Truncated exponential biphasic shocks (5/3 ms) were delivered at 20, 30, 50, 70 and 100 J. Four shocks at each energy level were delivered to construct energy vs. % success curves for VF termination. Electrode polarity for the first pulse was varied so that the first pulse cathode was either the apex (for apex-parasternal) or sternum (for sternum-vertebral column), or the reverse. The second pulse polarity was always the opposite of the first.
Results:
VF termination success rose from 0 to 86% as energy increased from 20 to 100 J. Varying the electrode polarity did not alter success rates at any energy level with either electrode pad placement.
Conclusion:
In this porcine model of transthoracic defibrillation, varying the biphasic shock electrode polarity did not alter transthoracic defibrillation success. Positional labeling of transthoracic biphasic defibrillation electrode pads may be unnecessary.