Recent progress on the vibrating flow pump as a totally implantable ventricular assist device
T Yambe1, S I Kobayashi, M Yoshizawa
1Department of Medical Engineering and Cardiology, Institute of Development, Aging and Cancer, Tohoku University, 4-1 Seiryo-machi, Aoba-ku, Sendai 980-77, Japan. yambe@idac.tohoku.ac.jp
This study details the vibrating flow pump (VFP) ventricular assist system, focusing on miniaturization for Asian patients. The VFP shows promise by increasing brain blood flow and influencing autonomic nerve function.
Area of Science:
- Biomedical Engineering
- Cardiovascular Devices
- Artificial Organs
Background:
- Development of smaller, lighter ventricular assist systems is crucial for patient populations like Asians.
- Current ventricular assist devices face challenges in size and weight for total implantation.
- Miniaturization of actuators is key for developing compact artificial hearts.
Purpose of the Study:
- To advance the development of a totally implantable vibrating flow pump (VFP) ventricular assist system.
- To achieve a smaller and lighter VFP system suitable for Asian patients by increasing drive frequency.
- To evaluate the physiological effects of VFP-assisted blood flow on circulation and autonomic function.
Main Methods:
- Utilized a cross slider form actuator for miniaturization, succeeding over linear and step motors.
- Designed blood-contacting parts, including a central vibration tube and round diaphragms, to prevent thrombus formation.
- Developed an energy transmission system for total implantation and analyzed autonomic nerve function using spectral analysis.
Main Results:
- Achieved significant miniaturization of the VFP system compared to previous actuator designs.
- Observed increased brain blood flow at frequencies of 25 to 30 Hz, without altering total circulatory flow.
- Detected observable influences on autonomic nerve function during VFP left heart assistance.
Conclusions:
- The vibrating flow pump (VFP) ventricular assist system shows progress towards a smaller, lighter, totally implantable device.
- The unique high-frequency oscillated blood flow pattern generated by the VFP impacts cerebral circulation and autonomic nervous system.
- Further research and development are ongoing with the goal of clinical application for VFP technology.