D A Devine1, A P Keech, D J Wood
1Oral Microbiology, Division of Oral Biology, Leeds Dental Institute, UK.
Aims:
To evaluate the antimicrobial efficacy of a novel u.v. beaker, powered in a domestic microwave oven.
Methods And Results:
Three beakers were compared, with most rapid killing obtained in the Neutra Plasma 50. Ultraviolet light generated within the beakers efficiently killed planktonic and surface-associated Streptococcus mutans, Pseudomonas aeruginosa, vegetative Bacillus stearothermophilus, herpes simplex and polio viruses. Candida albicans and Mycobacterium phleii were less rapidly killed, and only 70% inactivation of B. stearothermophilus endospores was achieved. Irradiation for 45 s reduced viable bacterial counts in saliva by > 99%.
Conclusions:
The u.v.-generating beakers efficiently reduced viable counts of bacteria, yeast and viruses. Kinetics of killing varied, reflecting the fact that lethal mechanisms are complex, and probably depend on interplay between u.v. and heat.
Significance And Impact Of The Study:
This novel method of generating u.v., using a cheap and widely available power source, provides a rapid, inexpensive and non-toxic method of disinfection with a wide range of applications in hospitals, clinics and the home.
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Temperature
Chemical Agents for Microbial Control
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...