"Tripolar" hip arthroplasty for failed hip resurfacing: nineteen years follow-up
1Department of Orthopedic Surgery and Traumatology Academisch Ziekenhuis van de Vrije Universiteit Brussel, Laarbeeklaan, 101, 1090 Brussels, Belgium.
Abstract:
The authors describe the case of a 37-year-old patient who sustained a subcapital femoral neck fracture six months after ICLH double-cup hip resurfacing. As the polyethylene acetabular resurfacing component was undamaged and well fixed, a standard femoral stem with a bipolar head was inserted. The outer diameter of the bipolar head was chosen to fit the resurfacing socket. The "tripolar" hip arthroplasty has functioned well for 19 years and was revised for aseptic cup loosening. The cemented femoral stem was still well fixed and was not revised. Although the "tripolar" hip has functioned well in our case, we believe it is not indicated for metal on metal bearings. In this case the use of an appropriate modular head with a correct head-socket clearance is preferred.