Abstract:
Ophthalmophakometric measurements of ocular surface radius of curvature and alignment were evaluated on physical model eyes encompassing a wide range of human ocular dimensions. The results indicated that defocus errors arising from imperfections in the ophthalmophakometer camera telecentricity and light source collimation were smaller than experimental errors. Reasonable estimates emerged for anterior lens surface radius of curvature (accuracy: 0.02-0.10 mm; precision 0.05-0.09 mm), posterior lens surface radius of curvature (accuracy: 0.10-0.55 mm; precision 0.06-0.20 mm), eye rotation (accuracy: 0.00-0.32 degrees; precision 0.06-0.25 degrees), lens tilt (accuracy: 0.00-0.33 degrees; precision 0.05-0.98 degrees) and lens decentration (accuracy: 0.00-0.07 mm; precision 0.00-0.07 mm).