Real-time pachymetry during photorefractive keratectomy using optical low-coherence reflectometry.
1University of Berne, Inselspital, Department of Ophthalmology, Berne, Switzerland.
Journal of Biomedical Optics
|December 1, 2001
Summary
Optical low-coherence reflectometry precisely measured central corneal thickness during photorefractive keratectomy. This technique accurately determined laser ablation depths, accounting for intraoperative corneal thinning.
Area of Science:
- Ophthalmology
- Biomedical Optics
- Surgical Technology
Background:
- Accurate intraoperative measurement of central corneal thickness is crucial for refractive surgery.
- Photorefractive keratectomy (PRK) involves laser ablation of corneal tissue, requiring precise depth control.
- Noncontact methods are desirable for intraoperative corneal assessment to minimize contamination and trauma.
Purpose of the Study:
- To evaluate optical low-coherence reflectometry (OLCR) as a noncontact method for measuring central corneal thickness during photorefractive keratectomy.
- To assess the feasibility of real-time OLCR measurements throughout the PRK procedure.
- To determine the accuracy of OLCR in calculating actual laser ablation depths, considering intraoperative corneal thinning.
Main Methods:
- Optical low-coherence reflectometry (OLCR) was employed for noncontact measurement of central corneal thickness.
- Continuous on-line measurements were taken intraoperatively: before surgery, after epithelium removal, during laser ablation, and post-ablation.
- Corneal thinning due to evaporation was quantified using OLCR to establish a baseline rate (-0.14 microm/s).
Main Results:
- OLCR successfully performed continuous, on-line measurements of central corneal thickness during PRK.
- A baseline corneal thinning rate of -0.14 microm/s was determined for the initial 5 minutes post-epithelium removal.
- Calculated ablation depths using OLCR showed a maximum difference of +/-10 microm compared to planned ablations (34-92 microm).
Conclusions:
- Optical low-coherence reflectometry is a viable noncontact tool for intraoperative central corneal thickness monitoring during PRK.
- Accounting for evaporative thinning improves the accuracy of measured laser ablation depths.
- OLCR offers potential for enhanced precision and safety in refractive surgical procedures.