Responses of a direct ion storage dosimeter (DIS-1) to heavy charged particles
1National Institute of Radiological Sciences, Anagawa 4-9-1, Inage, Chiba-shi 263-8555, Japan. h_yasuda@nirs.go.jp
Abstract:
The responses of a direct ion storage dosimeter (DIS-1) to energetic heavy charged particles were examined using (4)He, (12)C, (40)Ar and (56)Fe ion beams at the HIMAC at the National Institute of Radiological Sciences. The efficiency of the DIS-1 on the basis of absorbed dose was almost unity for the helium and carbon ions and was slightly decreased for the argon and iron ions. The linearity in the dose response and the angular independence for these heavy ions were fairly good. Although further studies are necessary, these results suggest that the DIS-1 would be a suitable passive dosimeter for measurements of absorbed dose in a field dominated by heavy charged particles such as the space environment.
Related Concept Videos
Biological Effects of Radiation
Determination of Crystal Structures
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides and...