Related Experiment Video
Updated: Jul 28, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Single capacitive discharge utilizing an auxiliary shock in the coronary venous system reduces the defibrillation
P R Roberts1, Y Zhang, S Zhuan
1Wessex Cardiothoracic Center, Southampton General Hospital, United Kingdom. DoctorPRR@aol.com
Unlabelled:
Auxiliary shocks (AS) from electrodes sutured to the left ventricle (LV) prior to primary biphasic shocks (PS) have been shown to reduce defibrillation thresholds (DFT). Two capacitors are required to generate these waveforms. We investigate delivery of AS from one capacitor using a novel waveform. The epicardial surface of the LV is accessed transvenously via the middle cardiac vein (MCV) avoiding a thoracotomy.
Methods:
A defibrillation electrode was placed in the right ventricle (RV) and superior vena cava (SVC) in 12 pigs (37+/-2 kg). A 50x1.8 mm electrode was inserted in the MCV through a guide catheter. A can was placed in the left pectoral region. A monophasic AS (100 microF, 1.5 J) was delivered along one pathway before switching to deliver a biphasic waveform (40% tilt, 2 ms phase 2) along another. DFTs (PS+AS) were assessed using a binary search. Two configurations not incorporating AS acted as controls. DFTs were compared using repeated measures analysis of variance.
Results:
DFTs of the four novel configurations (AS/PS) were: RV-->Can/MCV-->Can=14.9+/-3.7 J, MCV-->Can/RV-->Can=17.2+/-5.7 J, RV-->SVC+Can/MCV-->SVC+Can=13.4+/-4.6 J, MCV-->SVC+Can/RV-->SVC+Can=17.1+/-5.9 J. Delivering AS in the RV followed by PS in the MCV reduced the DFT (RV-->Can (19.9+/-7.3 J, P<0.01) and RV-->SVC+Can (19.2+/-6.0 J, P<0.05)).
Conclusions:
Delivering AS prior to PS in the MCV reduces the DFT by up to a third compared to conventional configurations of RV-->Can and RV-->SVC+Can. This is possible using only a single capacitor and an entirely transvenous approach to the LV.