False increase in serum C-reactive protein caused by monoclonal IgM-lambda: a case report.
Clinical Chemistry and Laboratory Medicine
|January 5, 2002
Summary
Human anti-animal antibodies can interfere with lab tests. A case study shows an IgM-lambda paraprotein falsely elevated C-reactive protein (CRP) levels, impacting diagnosis.
Area of Science:
- Clinical Chemistry
- Immunology
- Laboratory Medicine
Background:
- Human anti-animal antibodies are known interferents in laboratory diagnostics.
- High-titer autoantibodies can lead to inaccurate patient test results, affecting clinical decisions.
Observation:
- An 86-year-old man presented with markedly elevated C-reactive protein (CRP) levels (>500 mg/l).
- Discrepancies in CRP measurements across different assays prompted further investigation.
- Protein electrophoresis revealed a monoclonal IgM-lambda paraprotein that interfered with the initial CRP assay.
Findings:
- The patient's IgM-lambda paraprotein (7 g/l) specifically bound to goat anti-CRP antibodies used in an automated immunometric assay.
- Removal of the paraprotein corrected the falsely elevated CRP values, which then measured <12 mg/l.
- The paraprotein was attributed to monoclonal gammopathy of unknown significance, with no evidence of hematologic malignancy.
Implications:
- This case highlights the critical need to consider antibody-mediated interference in immunoassays.
- Accurate diagnostic interpretation requires awareness of paraproteins and their potential to cause false positives in C-reactive protein testing.
- Developing strategies to mitigate heterophile antibody interference is crucial for reliable laboratory diagnostics.