[The study of oxygen-conserving nasal cannula and evaluation of it's oxygen saving ratio in treatment]
1Department of Respiratory Diseases, China-Japan Friendship Hospital, Beijing 100029.
Objective:
To study the oxygen-conserving nasal cannula(CNC), and to evaluate it's oxygen saving benefit ratio in COPD patients.
Methods:
(1) CNC was designed as a device which conserved oxygen during exhalation and given oxygen during inhalation; (2) Evaluation of the effect of oxygen therapy: blood analysis was measured at different flows of oxygen (0.5, 1.0, 1.5, 2.0 L/min) before and after oxygen therapy. The test is randomized; Oxygen saturation(SPO2) was measured at 1, 2, 3, 5, 10, 15, 30, 60, 120 minutes by pulse oximeter; (3) Evaluation of the oxygen saving benefit ratio: SPO2 was measured at 0.5, 1.0, 1.5, 2.0, 2.5, 3.0 L/min using CNC, the standard nasal cannula (SCN) and the new CNC.
Results:
(1) During two hours of oxygen therapy, the values of PaO2 and SaO2 in the CNC group were always higher than those in the SNC group (P < 0.05); (2) When the supply flow to CNC was set at 0.5 L/min, 1.0 L/min, 1.5 L/min, 2.0 L/min seperately, the oxygen saving benefit ratio of CNC to SNC was 2.6:1, 1.8:1, 1.8:1 and 1.5:1 respectively. (3) The oxygen saving benefit ratio was comparable between CNC made in China and that made in foreign country.
Conclusions:
It was confirmed that CNC saved oxygen at different flows of oxygen, and the oxygen saving benefit ratio was higher at lower flows than that at higher flows.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: