Related Experiment Videos
Optical chamber system designed for microscopic observation of living cells under extremely high hydrostatic pressure
1Frontier Research System for Extremophiles, Japan Marine Science and Technology Center, Yokosuka. skoyama@jamstec.go.jp
Abstract:
A novel pressure chamber system has been developed for the study of living cells under conditions of extremely high hydrostatic pressure up to 100 MPa (1 atm = 0.101325 MPa). The temperature in the chamber is thermostatically controlled in the range from 2 degrees to 80 degrees C. Two high-pressure pumps are employed for continuous perfusion of the chamber with culture medium and a chemical solution under high hydrostatic pressure conditions. The chamber has a 2-mm-thick glass window 2 mm in diameter, with a minimum working distance of 3.8 mm. The chamber system is designed to be adaptable to a variety of microscopic and imaging techniques. Using this chamber system, we successfully carried out real-time observations of elongated Escherichia coli and rounded HeLa cells under pressure.