Potential rebreathing after continuous positive airway pressure failure during sleep
Ramon Farré1, Josep M Montserrat, Eugeni Ballester
1Unitat de Biofísica i Bioenginyeria, Facultat de Medicina, Universitat de Barcelona, Spain. farre@medicina.ub.es
Background:
Continuous positive airway pressure (CPAP) is widely applied as a home treatment during sleep. Conventional CPAP devices are based on a blower to generate nasal pressure and to maintain air washout from the circuit. Because common CPAP systems do not incorporate alarms, failure in the device or in the electric supply could result in rebreathing.
Aim:
To assess the potential rebreathing to which a patient could be subjected after CPAP failure.
Methods:
Four conventional CPAP devices, PV100 (Breas Medica; Molnlycke, Sweden), CP90 (Taema; Antony, France), and SoloPlus and BiPAP (Respironics, Murrysville, PA), and three common exhalation ports (Whisper Swivel [Respironics], Plateau [Respironics], and 4-mm orifice) were tested in a bench study. Rebreathing after failure was assessed by measuring the resistance of the exhalation port (REP) and the resistance of the tubing plus CPAP device (RTUB), and by measuring O(2) and CO(2) concentrations in the nasal mask in a subject breathing through a CPAP system.
Results:
REP was much higher (approximately 30 cm H(2)O x s/L) than RTUB (approximately 1 cm H(2)O x s/L). Most (approximately 90%) of the breathing tidal volume would flow from/to the tubing plus CPAP device, which represents a dead space (> or = 0.5 L) similar to the patient's tidal volume. After CPAP failure, end-tidal O(2) in the mask changed from 16.8 to 9.2% and end-tidal CO(2) in the mask changed from 4.2 to 6.2%. By contrast, O(2) and CO(2) did not change when a nonrebreathing valve was placed in the mask.
Conclusions:
Common CPAP systems run a risk of inducing rebreathing in case of failure. This risk could be easily avoided by including a passive valve in the apparatus.
More Related Videos
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
04:16Inspiratory Muscle Training as an Adjunct to the Treatment of Weaning Failure in Critically Ill Patients: A Practical Guide
Published on: January 30, 2026
Related Concept Videos
Acute Respiratory Failure-III
Acute Respiratory Failure-IV
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Sleep Apnea
The condition is more prevalent among...
Cardiopulmonary Resuscitation II: ACLS Airway Management