Related Experiment Video
Updated: Jul 1, 2026

Quantification of Heavy Metals and Other Inorganic Contaminants on the Productivity of Microalgae
Published on: July 10, 2015
Derivative hydride generation atomic absorption spectrometry and determination of lead traces in waters
1College of Chemistry and Environmental Science, Hebei University, Baoding, PR China. Han Wen@mail.hbu.edu.cn
Abstract:
Fundamentals of derivative hydride-generation atomic absorption spectrometry (DHGAAS) are described. A linear relationship was obtained between the derivative absorbance and the concentration of analysis in a sample. The new DHGAAS method was applied to the determination of traces of lead in water. The conditions affecting the derivative absorbance of lead were evaluated and optimized. The detection limit and sensitivity of the proposed method were 26 times and 8.8 times better, respectively, than those of conventional hydride-genera-tion atomic absorption spectrometry. The characteristic concentration (for a derivative absorbance of 0.0044) and detection limit (3sigma) for lead were 0.017 and 0.096 ng mL(-1), respectively, for a 2 mV min(-1) sensitivity range setting. The recovery range was 92.5-103%.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Lab