Related Experiment Video
Updated: Jun 29, 2026

Standardized Method for High-throughput Sterilization of Arabidopsis Seeds
Published on: October 17, 2017
Morpho-structural variations of bacterial spores after treatment in steam vacuum assisted autoclave
M Fonzi1, E Montomoli, R Gasparini
1Dipartimento di Scienze Biomediche, Università di Siena, Italia.
Abstract:
This study intended to verify, through microbiological techniques and TEM investigations, the killing of bacterial spores after treatment in steam autoclave, and to propose strictly morphological considerations about the target of this sterilisation process. Autoclave is the most common device for sterilising instruments in order to prevent cross infections in dental offices. The autoclave efficiency has been improved in the last years and part of this improvement is related to both a better and more correct use of the autoclave system and to the technological innovations introduced in the last generation of devices. However, associations as ADA or CDC suggest to regularly verify the process of 'autoclaving' through biological indicators (BI). The most commonly used BI are made of spores strips or suspensions of Bacillus Subtilis (pb 168) and Bacillus Stearothermophilus (ATCC 10149). They visually prove, changing colours on enzymatic base, the death of micro-organism and if the physical parameters, necessary for sterilisation, have been achieved. These two strains of endospore-forming bacteria were processed and prepared following two different techniques: Karnovsky fixed and epon embedded--phosphotungstic acid fixed for direct observation. The kind and the extent of analysed modifications are extremely various: from deep lacerations, which changed the spore structure, to little clefts which let the cytoplasm go out.
More Related Videos
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Special Staining Techniques
Microbial Morphologies
Endospores and Sporulation
Physical Methods for Controlling Microbial Growth: Temperature