The information content of Doppler ultrasound signals from the fetal heart
S A Shakespeare1, J A Crowe, B R Hayes-Gill
1School of Electrical and Electronic Engineering, University of Nottingham, UK.
Abstract:
Knowledge of the content of Doppler ultrasound signals from the fetal heart is essential if the performance of fetal heart rate (FHR) monitors based upon this technology is to be improved. For this reason instrumentation was constructed to enable the simultaneous collection of Doppler audio signals and the transabdominal fetal ECG (for signal registration), with a total of 22 recordings being made with an average length of around 20 minutes. These data demonstrate the transient nature of the Doppler audio data with wide variations in the signal content observable on a beat-to-beat basis. Short-time Fourier analysis enabled the content of the Doppler signals to be linked to six cardiac events, four valve and two wall motions, with higher frequency components being associated with the latter. This differing frequency content together with information regarding the direction of movement that can be discerned from Doppler signals provided a potential means of discriminating between these six events (which are unlikely to all contribute to the Doppler signal within the same cardiac cycle). Analysis of 100 records showed that wall contractions generate the most prominent signals, with atrial contraction recognisable in all records and ventricular wall contraction in 95% (although its amplitude is only around 30% of that of the atrial signal). Valve motion, with amplitudes between 15 and 25% that of the atrial wall signal, were visible in 75% of records. These results suggest means by which the six events that contribute to the Doppler signal may be distinguished, providing information that should enable an improvement in the current performance of Doppler ultrasound-based FHR monitors.
Related Concept Videos
Doppler Effect - II
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V) valves at the...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...