Inaccuracy of lipid measurements with the portable Cholestech L.D.X analyzer in patients with hypercholesterolemia
Related Experiment Video
Updated: Oct 2, 2026

Cell-free Biochemical Fluorometric Enzymatic Assay for High-throughput Measurement of Lipid Peroxidation in High Density Lipoprotein
Published on: October 12, 2017
James H Stein1, Cynthia M Carlsson, Kristi Papcke-Benson
1University of Wisconsin Medical School, Madison, WI 53792, USA.
Background:
Although total cholesterol concentrations measured by portable lipid analyzers have acceptable bias and precision in young and middle-aged adults, clinically relevant differences in HDL-cholesterol (HDL-C) and triglyceride values have been described. Furthermore, the accuracy of portable lipid analyzers in older hyperlipidemic individuals, who have a high incidence of coronary heart disease, has not been validated. This study determined the biases and variability in portable lipid measurements in older patients with hypercholesterolemia and related them to National Cholesterol Education Program Adult Treatment Panel III guidelines.
Methods:
Participants were > or =70 years of age with fasting serum LDL-cholesterol (LDL-C) concentrations > 1.40 g/L. Fasting fingerstick samples were analyzed on a Cholestech L.D.X desktop analyzer. Antecubital venous samples were analyzed in a proficiency-certified clinical laboratory.
Results:
Portable measurements systematically overestimated triglycerides (0.296 g/L; P <0.001) and HDL-C (0.015 g/L; P = 0.026). LDL-C concentrations were underestimated (0.043 g/L; P = 0.046). Total and non-HDL cholesterol calculations based on the portable lipid device provided unbiased estimates, but wide variability was present. Significant variability in lipid determinations limited their clinical usefulness in individual patients, especially because 2 SD of the mean bias between the laboratory and the portable determinations of LDL-C and non-HDL cholesterol exceeded the 0.30 g/L cutoff that defines treatment targets in the current lipid guidelines.
Conclusions:
Lipid values obtained from portable lipid analyzers may be useful for screening, but they should not be used to make clinical decisions regarding the diagnosis and management of dyslipidemia in individual patients.
Related Concept Videos
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test
Blood Studies for Cardiovascular System III: Serum Lipid Profile
Serum lipids are fats and fatty substances in the blood and are crucial for various bodily functions, including energy storage, cellular structure, and hormone production. Serum lipids consist of cholesterol, triglycerides, and phospholipids.
Cholesterol is a soft, fat-like substance found in all body cells. It is crucial for producing hormones, vitamin D, and substances that aid...