Cross-ventricular endless loop tachycardia during biventricular pacing.

S S Barold1, C L Byrd

  • 1Florida Cardiovascular Institute and Tampa General Hospital, USA. ssbarold@aol.com

Pacing and Clinical Electrophysiology : PACE
|January 31, 2002
PubMed
Summary

A novel pacemaker tachycardia occurred due to T-wave sensing in the atrial channel, triggering ventricular pacing. This "cross-ventricular" endless loop tachycardia can be prevented with proper pacemaker programming.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

True vs. pseudo-electrical ventricular alternans sustained by ventricular premature complexes.

Herzschrittmachertherapie & Elektrophysiologie·2018
Same author

A review of the atrial upper rate algorithms of St. Jude Medical (Abbott) cardiac implantable electronic devices : Incidence of repetitive nonreentrant ventriculoatrial synchrony (RNRVAS).

Herzschrittmachertherapie & Elektrophysiologie·2017
Same author

Erratum to: Noise from a dysfunctional atrial lead detected as atrial fibrillation by a cardiac implantable electronic device.

Herzschrittmachertherapie & Elektrophysiologie·2017
Same author

Diagnosis of myocardial infarction and ischemia in the setting of bundle branch block and cardiac pacing.

Herzschrittmachertherapie & Elektrophysiologie·2016
Same author

Unusual cause of desynchronization in a cardiac resynchronization device.

Herzschrittmachertherapie & Elektrophysiologie·2012
Same author

Atrial undersensing and cycle prolongation related to automatic mode switching: what is the mechanism?

Pacing and clinical electrophysiology : PACE·2012

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Electrophysiology

Background:

  • Conventional dual-chamber pacemakers are used for biventricular pacing.
  • VVIR mode is a common pacing setting.

Observation:

  • An unusual pacemaker tachycardia developed in a patient undergoing biventricular pacing.
  • The atrial channel was connected to the left ventricle, and the ventricular channel to the right ventricle.

Findings:

  • The tachycardia was sustained by T-wave sensing in the "atrial channel," which incorrectly triggered pacing in the "ventricular channel."
  • This resulted in a "cross-ventricular" endless loop tachycardia, a form of reentrant pacemaker tachycardia.

Implications:

  • This phenomenon highlights potential risks with unconventional lead placements in dual-chamber pacemakers.

Related Experiment Videos

  • Appropriate pacemaker programming is crucial for preventing such cross-ventricular endless loop tachycardias.
  • Understanding T-wave oversensing is vital for pacemaker safety and efficacy.