[Biomechanical studies and finite element analysis of a bone-implant interface]
Related Experiment Video
Updated: Oct 2, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
1Institut für Biomedizinische Technik und Krankenhaustechnik, Medizinische Hochschule Hannover. widjaja.wira@mh-hannover.de
Abstract:
In the present study, the fixation system of a femoral medullary nail connection was investigated. In surgical treatment of fractured femurs, the fracture is bridged by a medullary nail that is fixed by interlocking screws in the bone. Bone failure around these screws is the most common complication associated with the treatment of fractures of osteoporotic bone. The present study analyses the stresses present in the region of the implant/bone system. Three-dimensional finite element models were generated, a nonlinear structure analysis performed, and the stresses at material interfaces investigated. The highest concentration of stresses is to be found in the middle of the interlocking screws and the holes drilled in the bone. This is in agreement with the results of experimental investigations.