Neonatal magnetocardiography.

P G Anastasiadis1, P Anninos, A Kotini

  • 1Department of Obstetrics and Gynaecology, Medical School, Democritus University of Thrace, Alexandroupolis, Greece.

Clinical and Experimental Obstetrics & Gynecology
|February 13, 2002
PubMed
Summary

Magnetocardiography (MCG) effectively estimates neonatal cardiac rhythm in newborns. This non-invasive technique, using a single-channel SQUID, aids in diagnosing arrhythmias like ventricular tachycardia and bradycardia.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Association of maternal angiotensin II type 1 and type 2 receptor combination genotypes with susceptibility to early-onset preeclampsia.

Journal of human hypertension·2021
Same author

ANTIMÜLLERIAN HORMONE AS INDICATOR OF OVARIAN DYSFUNCTION.

Acta endocrinologica (Bucharest, Romania : 2005)·2019
Same author

Controversies on Transvaginal Ultrasound Screening for Endometrial Cancer in Asymptomatic Postmenopausal Women.

Revista medico-chirurgicala a Societatii de Medici si Naturalisti din Iasi·2018
Same author

Horner's syndrome and epidural anesthesia in labor and cesarean section.

Clinical and experimental obstetrics & gynecology·2018
Same author

Immunohistochemical expression of MTA1 and MTA3 in placental tissue of normal and preeclamptic pregnancies.

Clinical and experimental obstetrics & gynecology·2018
Same author

Correlation of age at natural menopause with occupational status and other epidemiologic factors in women from Prefecture of Kavala, Greece.

Hippokratia·2018

Area of Science:

  • Neonatal cardiology
  • Biomagnetism
  • Medical instrumentation

Background:

  • Neonatal cardiac rhythm assessment is crucial for early diagnosis of potential heart conditions.
  • Conventional methods may have limitations in speed and invasiveness.
  • Magnetocardiography (MCG) offers a potential non-invasive alternative.

Purpose of the Study:

  • To evaluate the accuracy of magnetocardiography (MCG) for assessing neonatal heart rhythms.
  • To determine the feasibility of using a single-channel superconductive quantum interference device (SQUID) for MCG in neonates.

Main Methods:

  • Fifty neonates from uncomplicated pregnancies and one with hypoplastic left heart syndrome were studied.
  • Magnetocardiography (MCG) was performed using a single-channel SQUID.

Related Experiment Videos

  • M-mode echocardiography was used for arrhythmia confirmation.
  • Main Results:

    • MCG identified normal cardiac rhythm in 44 neonates.
    • Arrhythmias detected included ventricular tachycardia (VT), VT with extrasystoles, and bradycardia.
    • The neonate with hypoplastic left heart syndrome showed ventricular bigeminy via MCG.

    Conclusions:

    • Magnetocardiography (MCG) is a valid, non-invasive method for evaluating neonatal cardiac rhythm.
    • MCG is rapid and easy to perform, serving as a valuable adjunct to traditional methods.
    • This technique can aid in the early detection and management of neonatal arrhythmias.