Automated construction of detector models for efficiency interpolation in gamma-ray spectrometry
Abstract:
A method was developed for automated construction of detector models in gamma-ray spectrometry, which can be used in Monte Carlo calculations of efficiency calibration curves. Full-energy peak efficiencies were first measured for different gamma-ray energies and for a given sample-detector arrangement and then calculated by the Monte Carlo method. For these calculations a detector model was employed along with a computer algorithm, which seeks agreement between the experimental efficiencies and the calculated ones by automatically determining the parameters of the model. The resulting agreement of the calculated data with the experimental one was within the relative uncertainty of the latter (3-4%) and the parameters of the detector models obtained were close to the values specified by the manufacturers. The detector models thus constructed can then be used for interpolation of calibration curves.