Continuous oxygen monitoring of the conjunctiva in neonates
Sherwin J Isenberg1, Doron Neumann, Stanley Fink
1Jules Stein Eye Institute, Departments of Ophthalmology and Pediatrics, Harbor/UCLA Medical Center, UCLA School of Medicine, Torrance, CA 90509, USA.
Objective:
To report the first use of a noninvasive conjunctival oxygen monitor in neonates. We wished to investigate if measuring the partial pressure of oxygen directly from the conjunctiva could track hyperoxia and hypoxia as well as, or better than, pulse oximetry. This has the potential to reflect brain oxygenation while yielding important systemic information.
Study Design:
Criteria standard.
Setting:
Tertiary care hospital neonatal intensive care unit.
Patients:
Ten newborns monitored with pulse oximetry.
Intervention:
The conjunctival oxygen monitor was studied for effectiveness and safety and compared to pulse oximetry.
Results:
The device remained on the eyes up to 120 minutes. The correlation coefficient between conjunctival oxygen tension and pulse oximetry was significant for all readings (p<0.001) and for pulse oximetry measurements below 100% (p=0.024). One infant developed eyelid edema, which subsided when the device was removed and one sustained a small corneal abrasion that healed overnight.
Conclusion:
This initial pilot report suggests that the conjunctival oxygen monitor is effective and relatively safe in neonates. The device holds promise to indicate a tissue oxygen reading, which may provide neonatologists with more information about systemic oxygenation.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: