Background:
Twenty-four-hour pH monitoring is an investigation technique that can give both false-positive and false-negative results, depending on patient factors such as diet and different activities. The aim was to study the reproducibility of 24-h oesophageal pH monitoring under as standardized conditions as possible in patients with symptoms of gastro-oesophageal reflux disease.
Methods:
Antimony pH electrodes were used in 22 adult patients who were investigated twice, 6 weeks apart, under identical conditions. They were hospitalized and were served a standardized diet which had been tested to contain no lower than pH 5.0. Neither coffee nor smoking was allowed. The patients had to refrain from proton-pump inhibitors for 10 days and H2 blockers for 2 days prior to the investigation. The paired t test was used.
Results:
There was no significant difference in total reflux time, upright or supine reflux time, or longest reflux periods between the two test occasions. However, there were discordant results in six patients who had normal total reflux time on one test occasion but pathological results on the other.
Conclusion:
Since the biological variability of gastro-oesophageal reflux is not negligible from time to time, a normal 24-h oesophageal pH test should be assessed with caution.
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: