Threshold-based fitting methods for non-linear (WDRC) hearing instruments--comparison of acoustic characteristics
1Department of Speech, Music, and Hearing, Royal Institute of Technology, Stockholm, Sweden. karolina@speech.kth.se
Comparing hearing aid fitting methods revealed significant variations in loudness perception, but minimal differences in speech intelligibility index (SII). Manufacturers
Area of Science:
- Audiology and Hearing Science
- Acoustics and Signal Processing
Background:
- Non-linear hearing instruments require precise prescriptive fitting methods for optimal performance.
- Existing prescriptive methods for hearing aids vary, potentially leading to different outcomes.
- Understanding the impact of different fitting strategies on audibility and speech intelligibility is crucial.
Purpose of the Study:
- To compare the performance of six threshold-based prescriptive fitting methods for non-linear hearing instruments.
- To evaluate the differences in estimated gain, calculated loudness, and speech intelligibility index (SII) across methods.
- To assess the accuracy of hearing aid manufacturers' implementation of selected prescriptive methods.
Main Methods:
- Six hearing aids were programmed using manufacturers' recommendations for a standard audiogram and simulated listening conditions.
- Coupler gain measurements were performed using speech-like signals.
- Loudness perception and Speech Intelligibility Index (SII) were calculated for each fitting method.
Main Results:
- Significant discrepancies were observed in the estimated insertion gain-frequency responses between the fitting methods.
- Calculated loudness showed substantial variations across methods, while SII calculations exhibited minimal differences.
- Large discrepancies were found between the prescribed gain and the actual implemented gain by hearing aid manufacturers for two methods (DSL[i/o] and FIG6).
Conclusions:
- Prescriptive fitting methods for non-linear hearing aids can lead to considerable differences in loudness perception.
- Speech intelligibility appears less sensitive to variations among these fitting methods compared to loudness.
- Hearing aid manufacturers' implementation of prescriptive formulas may deviate significantly from the original prescriptions, impacting device performance.