Kinetics of defibrillation shock-induced response: design implications for the optimal defibrillation waveform
K A Mowrey1, Y Cheng, P J Tchou
1Department of Cardiology, Cleveland Clinic Foundation, Ohio, USA.
Introduction:
Implantable cardioverter defibrillator (ICD) therapy is a well-established therapy for treating patients at high risk for sudden cardiac death. Recently formulated virtual electrode polarization theory is a promising foundation for the theory of defibrillation. Yet, continuing optimization of defibrillation therapy is limited to primarily empirical methods due to difficulties in assessing kinetics of cellular response in whole heart models of defibrillation. The aim of this study was to evaluate the response of the myocardium in the context of virtual electrode polarization.
Methods And Results:
We used a Langendorff-perfused rabbit heart model of ICD therapy and voltage-sensitive fluorescent dye imaging in order to map kinetics of trans membrane potential during both mono- and biphasic shocks applied at various phases of the QT-interval. Cellular response was fitted to a single exponential function using the Levenberg-Marquardt method. Time constants (tau) were measured in 45 288 optical records from 17 hearts. We found that cellular response depends upon both QT-phase of application, intensity, polarity, and phase of the biphasic waveform. Shocks of larger strengths produce a faster response. The tau of the first-phase negatively polarizing response was significantly larger compared with the positively polarizing response at intensities below 200 V, but smaller at 200 V and above. The tau of the second phase negatively polarizing response was always slower than the positively polarizing response, regardless of amplitude, and timing. Overall, tau ranged from 1.6 ms to 14.2 ms.
Conclusions:
The time constant of the membrane depends on the field, action potential phase and the shock polarity, but exceeds 1 msec. Therefore, we suggest using a slower shock leading edge, since the membrane cannot follow potentially damaging faster waveforms.
Related Concept Videos
Cardiopulmonary Resuscitation IV: Pharmacological Management
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrophysiology of Normal Cardiac Rhythm