Evaluation of the validity and reliability of A-scan ultrasound biometry with a single use disposable cover
Related Experiment Video
Updated: Oct 2, 2026

Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
K Cass1, C M Thompson, C Tromans
1Department of Optometry and Neuroscience, UMIST, PO Box 88, Manchester M60 1QD, UK. kevin.cass@stud.umist.ac.uk
Background:
The UK Medical Devices Agency has suggested that ophthalmic practitioners should, where practicable and not compromising clinical outcome, restrict corneal contact devices to single patient use to minimise a remote theoretical risk of transmission of new variant Creutzfeldt-Jakob disease (vCJD). This study reports on a modified technique of ultrasound A-scan biometry that complies with the MDA recommendations.
Methods:
The right eyes of 37 consecutive hospital patients had a series of biometry readings taken with a Humphrey 820 A-scan instrument with a plane wave transducer use d conventionally and with the addition of a disposable latex cover.
Results:
Intrasessional repeatability of axial length measurements was similar for conventional readings--mean difference 0.027 mm, 95% confidence intervals (CI) +/- 0.44 mm and those taken with a disposable cover (0.028 mm, CI +/- 0.38). Intersessional repeatability was equivalent with (0.002 mm, CI +.- 0.51) and without a cover (0.03 mm, CI +/- 0.51). Readings with a cover were not significantly different from those without (paired t test; p >0.05), but tended to be greater (mean difference 0.085 mm, CI +/- 0.60).
Conclusions:
These findings suggest that corneal contact biometry with a disposable cover is a viable and theoretically safer alternative to the conventional technique.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...