Related Experiment Video
Updated: Aug 18, 2026

Synthesis of Soft Polysiloxane-urea Elastomers for Intraocular Lens Application
Published on: March 8, 2019
[The choice of intraocular lens before cataract extraction for cases with high myopia]
1Institute of Ophthalmology, Shandong Academy of Medical Sciences, Qingdao 266071.
Objective:
To study the choice of diopter and size of optical portion of intraocular lens (IOL) before cataract extraction for cases with high myopia.
Methods:
Extracapsular cataract extraction (ECCE) or phacoemulsification with IOL implantation was performed on 60 cases (94 eyes) with high myopia (axial length > 26mm). The posterior chamber IOL selected and inserted was the one with the diopter and size of the optical portion suitable for the recipient to have a postoperative low myopia and easy to be examined with an ophthalmoscope.
Results:
The postoperative eyes were of low myopia, -2.07 +/- 1.95 D. The larger size of optical portion was more suitable for ocular fundus examination. Great care was taken for phacoemulsification for cataractous nucleus below III grade.
Conclusions:
For patients with axial myopia, the IOL implanted should create low myopia postoperatively. Larger optic size is better for fundus examination. Phacoemulsification can produce less astigmatism and damage, but the indication must be controlled strictly.
Related Concept Videos
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Focusing of Light in the Eye