Related Experiment Videos

Management of infected femoral closure devices

Kevin Geary1, Jeffrey T Landers, William Fiore

  • 1Rochester General Hospital Department of Vascular Surgery, 1445 Portland Avenue, Rochester, NY 14621, USA. kevin.geary@viahealth.org

Cardiovascular Surgery (London, England)
|March 13, 2002
PubMed

Insights

Infections and abscesses occurred in patients undergoing femoral artery closure with the Perclose device. Prompt drainage, suture removal, and surgical repair are recommended for these serious complications.

Area of Science:

  • Vascular Surgery
  • Infectious Diseases

Background:

  • Percutaneous femoral artery closure devices are increasingly used.
  • An increase in infectious complications has been observed with these devices.

Observation:

  • This report details five cases of infected groins and/or femoral arteries.
  • These infections occurred after angiographic procedures using the Perclose Suture Mediated Closure Device.

Findings:

  • All five patients developed abscesses requiring drainage.
  • Perclose suture removal was necessary in all cases.
  • Most patients needed extensive vascular reconstructive procedures.

Implications:

  • Infection is a significant risk associated with Perclose device use.
  • Timely intervention including abscess drainage, suture removal, and surgical repair is crucial.
  • Further research into preventing and managing these complications is warranted.

Related Concept Videos