Biomechanical evaluation of a double-threaded pedicle screw in elderly vertebrae
Related Experiment Video
Updated: Oct 2, 2026

An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Praveen V Mummaneni1, Sean M Haddock, Michael A K Liebschner
1Department of Neurological Surgery, University of California at San Francisco, 94143, USA.
Abstract:
We sought to test the hypothesis that a pedicle screw that has two parallel threads of different heights throughout the full length of the screw could increase both bone purchase and pull-out strength compared with a standard single-threaded screw of similar dimensions. A single-threaded pedicle screw and a double-threaded pedicle screw were respectively placed into the paired pedicles of 21 vertebral bodies. The screws were then pulled out of the pedicles, and output parameters were measured. Although insertional torque was, on average, 14.5% higher (p = 0.039) for the single-threaded screw, maximum pull-out strength (p = 0.12), energy-to-failure (p = 0.39), and stiffness (p = 0.54) were not statistically different for the two screw types. It is concluded that a second, smaller inner thread on a double-threaded pedicle screw does not translate into either increased bone purchase or higher pull-out strengths.