Intracavernosal Pressure Recording to Evaluate Erectile Function in Rodents
Published on: June 6, 2018
The correlation between penile tumescence measured by the erectometer and penile rigidity by the RigiScan
1Department of Urology, Sapporo Medical University, School of Medicine, Sapporo, Japan.
The erectometer accurately estimates penile rigidity, correlating well with RigiScan measurements. This tool is particularly useful for detecting significant changes in penile tumescence during nocturnal penile tumescence studies.
Area of Science:
- Urology
- Andrology
- Medical Devices
Background:
- Assessing penile rigidity is crucial for diagnosing erectile dysfunction.
- Nocturnal penile tumescence (NPT) studies are a standard diagnostic method.
- Comparing measurement devices for NPT is essential for clinical accuracy.
Purpose of the Study:
- To evaluate the correlation between penile tumescence measured by the erectometer and penile rigidity measured by the RigiScan.
- To determine the clinical utility of the erectometer in estimating penile rigidity patterns.
Main Methods:
- Nocturnal penile tumescence (NPT) was measured in 25 male volunteers.
- The erectometer and RigiScan were used simultaneously over 47 nights.
- Correlation between maximum penile circumferential changes (erectometer) and rigidity patterns (RigiScan) was analyzed.
Main Results:
- A strong correlation (r=0.719) was observed between penile circumferential changes measured by the erectometer and the RigiScan.
- Penile circumferential changes by erectometer corresponded well with RigiScan rigidity patterns (P=0.0001).
- Increments >30mm (70% PPV) and <10mm (100% NPV) by erectometer predicted normal rigidity patterns.
Conclusions:
- The erectometer shows good correlation with RigiScan for measuring penile rigidity.
- The erectometer is a valuable tool for estimating penile rigidity patterns, especially for significant circumferential changes.
- This suggests the erectometer can reliably assess penile rigidity during NPT studies.