A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Difficult tracheal intubation induced by maxillary distraction devices in craniosynostosis syndromes
John Roche1, Geoff Frawley, Andrew Heggie
1Great Ormond Street Hospital for Children, London, UK.
Background:
Difficult intubation occurred during anaesthesia for removal of maxillary distraction devices in five of seven children with syndromal craniosynostoses (four with Apert, two with Pfeiffer and one with Crouzon syndrome).
Methods:
Intubation was assessed in terms of laryngeal view and an established intubation difficulty score and had been straightforward before device insertion. Difficulty was induced by trismus due to device insertion and by increased maxillary prominence. This was compounded by preexisting mandibular hypoplasia. Cephalometric analysis, with each child acting as their own control, demonstrated anterior displacement of the maxilla and increased maxillary vertical height, as well as increased protuberance of the maxillary incisors.
Results:
All five difficult tracheal intubations were associated with preoperative Mallampati scores of 3 or 4 and the nine straightforward intubations with scores of 1 or 2. Maximal interincisor distance was less than the lower 95% confidence limit for age in all five children who were difficult to intubate at the time of device removal. No child had a failed intubation, but all had significantly increased intubation difficulty.
Conclusions:
In view of the risks of trauma, hypoxia and aspiration associated with difficult direct laryngoscopy, we recommend elective fibreoptic intubation at anaesthesia for removal of maxillary distraction osteogenesis devices in these children.
Related Concept Videos
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques