Portable electrochemical blood uric acid meter
Chin-Sung Kuo1, Chii-Min Hwu, Yueh-Hui Lin
1Endocrinology and Metabolism Section, Department of Medicine, Taipei Veterans General Hospital, Taiwan, Republic of China.
Journal of Clinical Laboratory Analysis
|April 12, 2002
Summary
The new UASure portable meter accurately measures uric acid (UA) in blood, showing good correlation with the standard Hitachi 7600 method. This device offers a reliable option for fast UA monitoring.
Area of Science:
- Biomedical Engineering
- Clinical Chemistry
- Medical Devices
Background:
- Hyperuricemia, a condition of elevated uric acid levels, is associated with gout and other health issues.
- Accurate and rapid monitoring of uric acid is crucial for managing hyperuricemic patients.
- Existing methods for uric acid measurement can be time-consuming or require laboratory facilities.
Purpose of the Study:
- To evaluate the performance of a new portable electrochemical uric acid meter, the UASure.
- To compare the accuracy and reliability of the UASure with a standard laboratory method (Hitachi 7600).
- To assess the suitability of the UASure for point-of-care uric acid testing.
Main Methods:
- The UASure meter was used to measure uric acid in capillary and venous blood samples from 146 volunteers.
- Uric acid concentrations were simultaneously measured using the standard Hitachi 7600 modular system.
- Intra-assay coefficients of variation and correlation coefficients (r) were calculated to assess precision and agreement.
Main Results:
- The UASure demonstrated low intra-assay coefficients of variation (3.08%–5.77%) across different uric acid levels.
- UASure measurements showed strong correlation with the Hitachi 7600 method (r = 0.87 for venous, r = 0.78 for capillary samples).
- The intraclass correlation for venous samples (rI = 0.84) indicated interchangeability with the standard method.
Conclusions:
- The portable UASure meter provides accurate and reliable uric acid measurements.
- UASure, particularly with venous blood samples, is interchangeable with the standard Hitachi 7600 method.
- The UASure is a promising device for rapid, point-of-care monitoring of uric acid concentrations.